Step-by-Step Guide to Setting Up Elevations Using Back-Sight on a Total Station

How to set up Topcon total station for leveling work only 

Here’s a detailed, step-by-step guide on using the back-sight method with a Total Station to set up elevations without needing to apply a resection coordinate system. This approach will help you obtain level data and marking levels.


Step 1: Initial Setup

  1. Power On the Total Station: Start by turning on the Total Station. Make sure the battery is fully charged for uninterrupted work.
  2. Select Job or Create New One: If your Total Station allows for different "jobs," choose the relevant job or create a new one for today’s measurements.
  3. Enter the Menu: Use the menu button to access the Total Station’s main functions.

Step 2: Accessing Back-Sight and Orientation Settings

  1. Choose Orientation Settings: In the menu, select the orientation option to start setting up the back-sight.
  2. Select Back-Sight Option: Within orientation, locate the back-sight function. This function lets you set a known point that the Total Station can reference for all other measurements.

Step 3: Setting the Back-Sight Point

  1. Choose Reception Mode: Inside the back-sight settings, go to reception mode. Reception mode lets you use reference points to determine the back-sight position.
  2. Select Elevation Requirement: Set the elevation setting to ensure you’re measuring levels accurately. This is essential if you’re only interested in getting elevation data.

Step 4: Targeting and Adjusting the Instrument

  1. Position the Instrument at the Target Point: Physically aim the Total Station at the target location where you want to set the back-sight.
  2. Set Target Height: Enter the target height based on the actual height of your setup. For example, you might input 0 if you’re measuring from ground level.
  3. Confirm Values: Press Yes or Enter to confirm the height value. The device will use this height as the reference for all other points.

Step 5: Checking Back-Sight and Orientation Data

  1. Return to Reception and Confirm Readings: In the reception section, check that the elevation reading reflects your desired setup.
  2. Observe Calibration Value: The Total Station should display a calibration or orientation value, such as 6.1. This value helps you ensure that the instrument is correctly oriented.
  3. Adjust If Necessary: If there’s an error message (e.g., "bad condition"), recheck your setup, adjust the instrument, and confirm again until the calibration is stable.

Step 6: Verifying with a Known Point

  1. Re-enter Coordinator and Observation: To confirm your setup, return to the coordinator and observation sections in the menu.
  2. Set Observation Point: Use a known point for observation to verify that the calibration and orientation match expected values. This will confirm that your setup is accurate.

Step 7: Final Confirmation

  1. Review Elevation and Calibration: Check the calibration and orientation values one last time to confirm consistency.
  2. Complete the Setup: Once all values match up correctly, you’re ready to start taking measurements.

Conclusion

  • Save Settings: Ensure your settings and calibration are saved for future use.
  • Shutdown or Proceed with Measurements: If you’re finished, shut down the device. If you need more measurements, proceed to collect data at different points.

Closing Notes

Thank your viewers and encourage them to subscribe for more detailed guides and tutorials!

Post a Comment

0 Comments